ASME V Article 4

Rexolite wedge on carbon steel

A 36° incidence in Rexolite is proposed for a shear-wave exam in carbon steel.

Reading the result. The angle is between the two critical angles. The refracted shear angle is about 54.33° from the normal.

Setup

Wedge or couplant

°
Measured from the normal.

Test material

Worked examples

Results update as you edit. Use calibrated measurements.

Inspection readout

SI units
Refracted L
—
Refracted S
54.33°
Critical 1
23.251°
Critical 2
46.347°
Pure shear-wave window

L is past its critical angle. The shear mode can propagate into the part.

Refraction at the wedge interfaceAll angles are measured from the surface normal. The lower band marks the kinematic regimes and current incidence.Wedge interfaceFrom normalIncidence / critical bands36°Incident rayL modeS mode
All angles are measured from the surface normal. The lower band marks the kinematic regimes and current incidence.
Wedge L velocity used
2337m/s
Part L velocity used
5920m/s
Part S velocity used
3230m/s
Wedge acoustic impedance
2.4539MRayl
Part acoustic impedance
46.472MRayl
Normal power reflection
80.944%
Normal power transmission
19.056%
Normal one-way interface loss
7.1998dB

The reflection, transmission and loss figures apply to longitudinal waves at normal incidence. Material presets are typical values. Calibrate beam index and angle on a suitable reference block. Equations and references