10,000 × g to RPM at an 85 mm Radius
Convert 10,000 × g to rpm for an 85 mm maximum microcentrifuge rotor radius and see the force along the tube.
RCF = 1.118e−5 × r_cm × RPM²
Bench result
Calculated in this browser
- Calculate
- Speed from RCF
- Reference radius
- r_max (tube bottom)
- Target RCF (× g)
- 10000
- Radius unit
- mm
- Maximum radius
- 85
- Minimum radius
- 65
- Speed
- 10,258.2rpm
- RCF at r_max
- 10,000× g
- Rotor k factor
- 645.748
| Position | Radiusmm | RCF× g |
|---|---|---|
| r_min | 65 | 7,647.06 |
| r_avg | 75 | 8,823.53 |
| r_max | 85 | 10,000 |
Worked calculation
Convert reference radius to centimeters
r_max = 85 mm = 8.5 cm.
Solve for the speed
RPM = √(10,000 ÷ (1.118e−5 × 8.5)) = 10,258.2 rpm.
Estimate the k factor
k = 10¹³ × ln(85 ÷ 65) ÷ (3600 × ω²) = 645.748 at ω = 1,074.23 rad/s.
- Check the instrument, rotor and tube speed ratings. The k-factor estimate uses the entered geometry and speed. Compare runs at the same sedimentation conditions.
How the model works
Enter the distance from the rotation axis to the relevant point in the tube. Use the same radius definition as the protocol or instrument specification.
Frequently asked questions
What is 85 mm in the RCF formula?
The formula with a coefficient of 1.118e−5 uses centimeters, so enter 8.5 cm in the equation.
Which speed should I choose if my display uses whole RPM?
The mathematical result is about 10,258 rpm. Follow the protocol tolerance and the instrument’s allowed increments and ratings.