Centrifuge RCF to RPM Converter

Convert × g and rpm at the minimum, midpoint or maximum rotor radius. Estimate the k factor and scale pelleting time.

RCF = 1.118e−5 × r_cm × RPM²

Inputs

Optional for speed or RCF alone
Blank uses the midpoint
Optional
Optional with reference time
Optional with reference k factor

Bench result

Calculated in this browser

Speed
10,258.2rpm
RCF at r_max
10,000× g
Rotor k factor
645.748
RCF across rotor radius7,6508,2408,8209,41010,0006570758085RCF (× g)Rotor radius (mm)r_max (85, 10,000)RCF at 10,258.2 rpm
RCF across rotor radius
RCF by radial position
PositionRadiusmmRCF× g
r_min657,647.06
r_avg758,823.53
r_max8510,000

How the model works

Enter the distance from the rotation axis to the relevant point in the tube. Use the same radius definition as the protocol or instrument specification.

Frequently asked questions

Which rotor radius is used for a maximum RCF specification?

Use the distance to the farthest point in the filled tube, usually near the bottom. Confirm the rated geometry in the instrument documentation.

How does rotor k factor affect pelleting time?

At the same sedimentation coefficient, time is proportional to k. Use k factors at the actual run speeds and comparable solvent conditions.

Related worked examples