Convert × g and rpm at the minimum, midpoint or maximum rotor radius. Estimate the k factor and scale pelleting time.
RCF = 1.118e−5 × r_cm × RPM²
Bench result
Calculated in this browser
- Calculate
- Speed from RCF
- Reference radius
- r_max (tube bottom)
- Target RCF (× g)
- 10000
- Radius unit
- mm
- Maximum radius
- 85
- Minimum radius
- 65
- Speed
- 10,258.2rpm
- RCF at r_max
- 10,000× g
- Rotor k factor
- 645.748
| Position | Radiusmm | RCF× g |
|---|---|---|
| r_min | 65 | 7,647.06 |
| r_avg | 75 | 8,823.53 |
| r_max | 85 | 10,000 |
Calculation steps
Convert reference radius to centimeters
r_max = 85 mm = 8.5 cm.
Solve for the speed
RPM = √(10,000 ÷ (1.118e−5 × 8.5)) = 10,258.2 rpm.
Estimate the k factor
k = 10¹³ × ln(85 ÷ 65) ÷ (3600 × ω²) = 645.748 at ω = 1,074.23 rad/s.
- Check the instrument, rotor and tube speed ratings. The k-factor estimate uses the entered geometry and speed. Compare runs at the same sedimentation conditions.
How the model works
Enter the distance from the rotation axis to the relevant point in the tube. Use the same radius definition as the protocol or instrument specification.
Frequently asked questions
Which rotor radius is used for a maximum RCF specification?
Use the distance to the farthest point in the filled tube, usually near the bottom. Confirm the rated geometry in the instrument documentation.
How does rotor k factor affect pelleting time?
At the same sedimentation coefficient, time is proportional to k. Use k factors at the actual run speeds and comparable solvent conditions.