EV housing CMM study
AIAG MSA · 4th edition / SI units
- GRR sigma
- 2.038489 µm
- GRR study var
- 7.9858 %
- GRR tolerance
- 10.192 %
- Distinct groups
- 17
Inputs
- Input mode
- Summary ranges
- Verdict basis
- % study variation
- Appraisers
- 2
- Trials per part
- 3
- Parts
- 7
- LSL
- 39.94 mm
- USL
- 40.06 mm
- Average repeat range
- 3.3 µm
- Appraiser mean range
- 1.1 µm
- Part mean range
- 72 µm
Measurement system components
| Component | Sigma | Study var. | Contribution | Tolerance |
|---|---|---|---|---|
| EV | 1.930795 µm | 7.5639 % | 0.57213 % | 9.654 % |
| AV | 0.6538088 µm | 2.5613 % | 0.065603 % | 3.269 % |
| GRR | 2.038489 µm | 7.9858 % | 0.63773 % | 10.192 % |
| PV | 25.4448 µm | 99.681 % | 99.362 % | 127.22 % |
| TV | 25.52633 µm | 100 % | 100 % | 127.63 % |
Range checks
| Check | Result |
|---|---|
| R̄̄ · average repeat range | 3.3 µm |
| X̄diff · appraiser mean range | 1.1 µm |
| Rp · part mean range | 72 µm |
| K₁ / K₂ / K₃ | 0.58509 / 0.7071 / 0.3534 |
| ndc ≥ 5 | Met |
Verdict based on % study variation. ndc meets the screening goal. Use representative parts and blinded, randomized repeats. The method does not estimate appraiser × part interaction. A zero observed GRR gives no finite reliability estimate for ndc.