Optical fixture pin study
AIAG MSA · 4th edition / SI units
- GRR sigma
- 1.329649 µm
- GRR study var
- 48.185 %
- GRR tolerance
- 30.684 %
- Distinct groups
- 2
Inputs
- Input mode
- Raw crossed study
- Verdict basis
- % drawing tolerance
- Appraisers
- 3
- Trials per part
- 3
- Parts
- 5
- LSL
- 2.987 mm
- USL
- 3.013 mm
Raw readings (mm), grouped by appraiser
- Part 1
- 2.9915 / 2.9923 / 2.9931 · 2.9941 / 2.9921 / 2.9929 · 2.9939 / 2.9947 / 2.9927
- Part 2
- 2.9929 / 2.9941 / 2.9953 · 2.9955 / 2.9939 / 2.9951 · 2.9953 / 2.9965 / 2.9949
- Part 3
- 2.9943 / 2.9959 / 2.9975 · 2.9969 / 2.9957 / 2.9973 · 2.9967 / 2.9983 / 2.9971
- Part 4
- 2.9957 / 2.9965 / 2.9973 · 2.9983 / 2.9963 / 2.9971 · 2.9981 / 2.9989 / 2.9969
- Part 5
- 2.9971 / 2.9983 / 2.9995 · 2.9997 / 2.9981 / 2.9993 · 2.9995 / 3.0007 / 2.9991
Measurement system components
| Component | Sigma | Study var. | Contribution | Tolerance |
|---|---|---|---|---|
| EV | 1.124095 µm | 40.736 % | 16.594 % | 25.941 % |
| AV | 0.710195 µm | 25.737 % | 6.6237 % | 16.389 % |
| GRR | 1.329649 µm | 48.185 % | 23.218 % | 30.684 % |
| PV | 2.418 µm | 87.625 % | 76.782 % | 55.8 % |
| TV | 2.759473 µm | 100 % | 100 % | 63.68 % |
Range checks
| Check | Result |
|---|---|
| R̄̄ · average repeat range | 1.92 µm |
| X̄diff · appraiser mean range | 1.466667 µm |
| Rp · part mean range | 6 µm |
| K₁ / K₂ / K₃ | 0.58547 / 0.5231 / 0.403 |
| ndc ≥ 5 | Below goal |
| R upper control limit | 4.94208 µm |
| X̄ control interval | 2.994189 mm to 2.998117 mm |
| R above UCL | 0 / 15 |
| X̄ outside control limits | 7 / 15 |
Verdict based on % drawing tolerance. Discrimination needs review. Use representative parts and blinded, randomized repeats. The method does not estimate appraiser × part interaction. A zero observed GRR gives no finite reliability estimate for ndc.