Laser stage height gage
AIAG MSA · 4th edition / SI units
- GRR sigma
- 4.15934 µm
- GRR study var
- 80.472 %
- GRR tolerance
- 124.78 %
- Distinct groups
- 1
Inputs
- Input mode
- Summary ranges
- Verdict basis
- % drawing tolerance
- Appraisers
- 2
- Trials per part
- 2
- Parts
- 6
- LSL
- 19.99 mm
- USL
- 20.01 mm
- Average repeat range
- 2.5 µm
- Appraiser mean range
- 5.1 µm
- Part mean range
- 8.2 µm
Measurement system components
| Component | Sigma | Study var. | Contribution | Tolerance |
|---|---|---|---|---|
| EV | 2.164682 µm | 41.881 % | 17.54 % | 64.94 % |
| AV | 3.551656 µm | 68.715 % | 47.217 % | 106.55 % |
| GRR | 4.15934 µm | 80.472 % | 64.757 % | 124.78 % |
| PV | 3.06844 µm | 59.366 % | 35.243 % | 92.053 % |
| TV | 5.168698 µm | 100 % | 100 % | 155.06 % |
Range checks
| Check | Result |
|---|---|
| R̄̄ · average repeat range | 2.5 µm |
| X̄diff · appraiser mean range | 5.1 µm |
| Rp · part mean range | 8.2 µm |
| K₁ / K₂ / K₃ | 0.86587 / 0.7071 / 0.3742 |
| ndc ≥ 5 | Below goal |
Verdict based on % drawing tolerance. Discrimination needs review. Use representative parts and blinded, randomized repeats. The method does not estimate appraiser × part interaction. A zero observed GRR gives no finite reliability estimate for ndc.