Pin gage limited repeats
JCGM 100 / ASME B89.7.3.1 / SI units
- Combined uᶜ
- 1.404251 µm
- Expanded U
- 5.238851 µm
- Coverage k
- 3.73071
- TUR
- 1.14529
Inputs
- Measured result
- 3.0012 mm
- Reference length
- 3 mm
- Lower specification
- 2.994 mm
- Upper specification
- 3.006 mm
- Decision rule
- Guard band · ±U
- Coverage factor
- Student-t · 95.45%
- Repeatability SD
- 2.3 µm
- Repeat count
- 3
- Instrument input
- MPE · rectangular half-width
- Instrument uncertainty / MPE
- 0.6 µm
- Instrument df (optional)
- —
- Resolution increment
- 0.25 µm
- Probe / form half-width
- 0.4 µm
- Fixture / alignment half-width
- 0.3 µm
- Workpiece temperature
- 22.5 °C
- Estimated differential CTE
- 2.7 ppm/°C
- Differential CTE uncertainty
- 1.4 ppm/°C
- Temperature uncertainty
- 0.6 °C
- Thermal input distributions
- Rectangular half-widths
- Apply nominal thermal correction
- No
Uncertainty budget
| Component | Standard u | Degrees of freedom | Variance share |
|---|---|---|---|
| Repeatability | 1.327906 µm | 2 | 89.422 % |
| Instrument | 0.3464102 µm | ∞ | 6.0854 % |
| Resolution | 0.07216878 µm | ∞ | 0.26412 % |
| Probe / form | 0.2309401 µm | ∞ | 2.7046 % |
| Fixture / alignment | 0.1732051 µm | ∞ | 1.5214 % |
| Thermal mismatch | 0.006732667 µm | ∞ | 0.0022987 % |
Conformance and coverage
| Check | Result |
|---|---|
| Measured result used | 3.0012 mm |
| Applied thermal correction | 0 µm |
| Reported interval | 2.995961149 mm to 3.006438851 mm |
| Guarded acceptance zone | 2.999238851 mm to 3.000761149 mm |
| Simple acceptance | Accepted |
| Guarded result | Indeterminate |
| Closest decision margin | −0.438851 µm |
| Effective degrees of freedom | 2.50115 |
| Coverage basis | 95.45 % two-sided |
Decision based on the measured interval, including expanded uncertainty. The illustrated three-zone guard rule accepts values whose full ±U interval lies inside spec, rejects those wholly outside, and flags overlap. A contract may specify another rule, correlations, temperature correction or uncertainty coverage.