Optics bench comparator
JCGM 100 / ASME B89.7.3.1 / SI units
- Combined uᶜ
- 2.160447 µm
- Expanded U
- 4.551947 µm
- Coverage k
- 2.10695
- TUR
- 4.39372
Inputs
- Measured result
- 75.021 mm
- Reference length
- 75 mm
- Lower specification
- 74.98 mm
- Upper specification
- 75.02 mm
- Decision rule
- Simple acceptance
- Coverage factor
- Student-t · 95.45%
- Repeatability SD
- 1.5 µm
- Repeat count
- 8
- Instrument input
- Expanded certificate uncertainty
- Instrument uncertainty / MPE
- 3.6 µm
- Certificate k
- 2
- Instrument df (optional)
- 12
- Resolution increment
- 1 µm
- Probe / form half-width
- 1.4 µm
- Fixture / alignment half-width
- 1.1 µm
- Workpiece temperature
- 19.5 °C
- Estimated differential CTE
- 2.7 ppm/°C
- Differential CTE uncertainty
- 1.4 ppm/°C
- Temperature uncertainty
- 0.6 °C
- Thermal input distributions
- Rectangular half-widths
- Apply nominal thermal correction
- No
Uncertainty budget
| Component | Standard u | Degrees of freedom | Variance share |
|---|---|---|---|
| Repeatability | 0.5303301 µm | 7 | 6.0257 % |
| Instrument | 1.8 µm | 12 | 69.416 % |
| Resolution | 0.2886751 µm | ∞ | 1.7854 % |
| Probe / form | 0.8082904 µm | ∞ | 13.997 % |
| Fixture / alignment | 0.6350853 µm | ∞ | 8.6413 % |
| Thermal mismatch | 0.07924961 µm | ∞ | 0.13456 % |
Conformance and coverage
| Check | Result |
|---|---|
| Measured result used | 75.021 mm |
| Applied thermal correction | 0 µm |
| Reported interval | 75.016448053 mm to 75.025551947 mm |
| Guarded acceptance zone | 74.984551947 mm to 75.015448053 mm |
| Simple acceptance | Rejected |
| Guarded result | Indeterminate |
| Closest decision margin | −1 µm |
| Effective degrees of freedom | 24.5862 |
| Coverage basis | 95.45 % two-sided |
Decision based on the reported value relative to drawing limits. The illustrated three-zone guard rule accepts values whose full ±U interval lies inside spec, rejects those wholly outside, and flags overlap. A contract may specify another rule, correlations, temperature correction or uncertainty coverage.