Chamber pump-down time / Worked example

Humid Chamber Recovery

A 460 L chamber has 4.8 m² of newly exposed walls. Use a slow wall decay to explore recovery after humid venting.

Readout ↓

Inputs

Chamber and effective speeds
L
Pa
Pa
Pa
L/s
L/s
Roots booster
Walls, seals and leaks
m²
Pa·m³/(s·m²)
s
Age at the start of the high-vacuum stage.
Pa·m³/(s·m²)
Pa·m³/s

Live readout

Target vacuum met
Roughing time218sHigh-vac time2115sTotal time2333sUltimate floor1.41e-7PaS₁Backing218 sS₂Roots bypass0 sS₃High vacuum2110 sLog P · Pa vs log-spaced time (s)1.41e-70.121.01e5075.22730Roots cut-inInactiveTarget P4e-4 PaFloor P1.41e-7 Pa
Roughing time
218 s
Backing-only roughing time
218 s
Boosted roughing time
0 s
High-vacuum stage
2115 s
Time to target
2333 s
Modeled ultimate pressure
1.41e-7 Pa
Wall throughput at target
8.394e-5 Pa·m³/s

    Reading this example

    The pressure can fall far below the crossover quickly while still requiring much longer to reach the final target.

    Model & assumptions

    Roughing stages

    t = (V/S) ln(Pₐ/Pᵦ)

    The backing pump runs above the Roots switch pressure. Use combined speed below the switch, until the high-vacuum pump takes over. Enter chamber-effective speeds for all stages.

    High-vacuum mass balance

    V dP/dt = A[q∞ + q₁(1 + t/t₁)⁻ᵅ] + Qₗ − SₑP

    The high-vacuum clock starts with wall age t₁. The solver integrates the decaying load with exponential chamber memory until pressure crosses the target. At α = 0 the reference wall rate is a steady load.

    Ultimate pressure

    P∞ = (Aq∞ + Qₗ)/Sₑ (α > 0)

    The model rejects a target at or below the ultimate pressure. With a decaying wall load, a target can sit above the floor and still take longer than the 10¹² s search horizon.