Helium leak & response time / Worked example
10% Helium Bag Test for a Wafer Module
A wafer module is bagged with 10% helium. The detector reads 4 × 10⁻¹⁰ Pa·m³/s with a 38 L test volume.
Inputs
Rate-of-rise isolation test
Live readout
Molecular helium trace- 100% helium equivalent
- 4e-9 Pa·m³/s
- Gas equivalence multiplier
- 0.3717
- Equivalent gas throughput
- 1.487e-9 Pa·m³/s
- 63.2% response (1τ)
- 5.067 s
- 95.0% response (3τ)
- 15.2 s
- 99.3% response (5τ)
- 25.33 s
- Isolation gas load
- 2.217e-7 Pa·m³/s
Reading this example
Correct for tracer fraction before converting to air-equivalent throughput. Allow 3τ for about 95% of the modeled signal.
Model & assumptions
Tracer correction
Qgas = (QHe/fHe) × √(MHe/Mgas) (molecular)
For a laminar viscous leak path use ηHe/ηgas with representative viscosities at 20 °C. These ratios assume identical geometry, pressure conditions and temperature.
Detector and isolation
τ = V/SHe · Signal = 1 − e⁻ᵗ/ᵗᵃᵘ · Qrise = V ΔP/Δt
The signal curve assumes a well-mixed volume and first-order response. An isolated chamber can rise from desorption or permeation as well as an external leak.