Helium leak & response time / Worked example
R-134a Vapor Leak Equivalence
A refrigeration component is screened with helium before charging. Estimate an R-134a vapor equivalent under effusive conditions.
Inputs
Rate-of-rise isolation test
Live readout
Molecular helium trace- 100% helium equivalent
- 2.8e-9 Pa·m³/s
- Gas equivalence multiplier
- 0.1981
- Equivalent gas throughput
- 5.546e-10 Pa·m³/s
- 63.2% response (1τ)
- 5.938 s
- 95.0% response (3τ)
- 17.81 s
- 99.3% response (5τ)
- 29.69 s
- Isolation gas load
- 5.542e-8 Pa·m³/s
Reading this example
Use the calculated value as a gas-phase comparison. Service pressure, liquid flow and production limits need their own test basis.
Model & assumptions
Tracer correction
Qgas = (QHe/fHe) × √(MHe/Mgas) (molecular)
For a laminar viscous leak path use ηHe/ηgas with representative viscosities at 20 °C. These ratios assume identical geometry, pressure conditions and temperature.
Detector and isolation
τ = V/SHe · Signal = 1 − e⁻ᵗ/ᵗᵃᵘ · Qrise = V ΔP/Δt
The signal curve assumes a well-mixed volume and first-order response. An isolated chamber can rise from desorption or permeation as well as an external leak.