Helium leak & response time / Worked example

Water Vapor Leak Equivalence

A vacuum drying vessel uses a 40% tracer mix and estimates a water-vapor-equivalent molecular rate.

Readout ↓

Inputs

Tracer and detector
Pa·m³/s
%
L
L/s
Rate-of-rise isolation test
Pa
s

Live readout

Molecular helium trace
Equivalent Q1.414e-9Pa·m³/s1τ response24.44s3τ response73.33sRate-of-rise Q3.872e-7Pa·m³/sHeTracer mix40%VTest volume220 LSDetector9 L/sDetector signal · % of final response0%50%100%01τ3τ5τ1τ · 63.2%24.4 s3τ · 95.0%73.3 s5τ · 99.3%122 s
100% helium equivalent
3e-9 Pa·m³/s
Gas equivalence multiplier
0.4714
Equivalent gas throughput
1.414e-9 Pa·m³/s
63.2% response (1τ)
24.44 s
95.0% response (3τ)
73.33 s
99.3% response (5τ)
122.2 s
Isolation gas load
3.872e-7 Pa·m³/s

    Reading this example

    Water adsorption and condensation can slow or distort an actual response. The equivalent rate here uses molecular mass only.

    Model & assumptions

    Tracer correction

    Qgas = (QHe/fHe) × √(MHe/Mgas) (molecular)

    For a laminar viscous leak path use ηHe/ηgas with representative viscosities at 20 °C. These ratios assume identical geometry, pressure conditions and temperature.

    Detector and isolation

    τ = V/SHe · Signal = 1 − e⁻ᵗ/ᵗᵃᵘ · Qrise = V ΔP/Δt

    The signal curve assumes a well-mixed volume and first-order response. An isolated chamber can rise from desorption or permeation as well as an external leak.