Helium leak & response time / Worked example

R-134a Vapor Leak Equivalence

A refrigeration component is screened with helium before charging. Estimate an R-134a vapor equivalent under effusive conditions.

Readout ↓

Inputs

Tracer and detector
Pa·m³/s
%
L
L/s
Rate-of-rise isolation test
Pa
s

Live readout

Molecular helium trace
Equivalent Q5.546e-10Pa·m³/s1τ response5.938s3τ response17.81sRate-of-rise Q5.542e-8Pa·m³/sHeTracer mix25%VTest volume9.5 LSDetector1.6 L/sDetector signal · % of final response0%50%100%01τ3τ5τ1τ · 63.2%5.94 s3τ · 95.0%17.8 s5τ · 99.3%29.7 s
100% helium equivalent
2.8e-9 Pa·m³/s
Gas equivalence multiplier
0.1981
Equivalent gas throughput
5.546e-10 Pa·m³/s
63.2% response (1τ)
5.938 s
95.0% response (3τ)
17.81 s
99.3% response (5τ)
29.69 s
Isolation gas load
5.542e-8 Pa·m³/s

    Reading this example

    Use the calculated value as a gas-phase comparison. Service pressure, liquid flow and production limits need their own test basis.

    Model & assumptions

    Tracer correction

    Qgas = (QHe/fHe) × √(MHe/Mgas) (molecular)

    For a laminar viscous leak path use ηHe/ηgas with representative viscosities at 20 °C. These ratios assume identical geometry, pressure conditions and temperature.

    Detector and isolation

    τ = V/SHe · Signal = 1 − e⁻ᵗ/ᵗᵃᵘ · Qrise = V ΔP/Δt

    The signal curve assumes a well-mixed volume and first-order response. An isolated chamber can rise from desorption or permeation as well as an external leak.