Helium leak & response time / Worked example
Water Vapor Leak Equivalence
A vacuum drying vessel uses a 40% tracer mix and estimates a water-vapor-equivalent molecular rate.
Inputs
Rate-of-rise isolation test
Live readout
Molecular helium trace- 100% helium equivalent
- 3e-9 Pa·m³/s
- Gas equivalence multiplier
- 0.4714
- Equivalent gas throughput
- 1.414e-9 Pa·m³/s
- 63.2% response (1τ)
- 24.44 s
- 95.0% response (3τ)
- 73.33 s
- 99.3% response (5τ)
- 122.2 s
- Isolation gas load
- 3.872e-7 Pa·m³/s
Reading this example
Water adsorption and condensation can slow or distort an actual response. The equivalent rate here uses molecular mass only.
Model & assumptions
Tracer correction
Qgas = (QHe/fHe) × √(MHe/Mgas) (molecular)
For a laminar viscous leak path use ηHe/ηgas with representative viscosities at 20 °C. These ratios assume identical geometry, pressure conditions and temperature.
Detector and isolation
τ = V/SHe · Signal = 1 − e⁻ᵗ/ᵗᵃᵘ · Qrise = V ΔP/Δt
The signal curve assumes a well-mixed volume and first-order response. An isolated chamber can rise from desorption or permeation as well as an external leak.