Flange load & viewport stress / Worked example
Clamped 304 Steel Blank Cover
A 180 mm steel blank plate closes a high-vacuum vessel with a 220 mm seal diameter.
Inputs
Seal and gland
Live readout
Viewport and seal screen pass- Pressure differential
- 1.013e5 Pa
- Atmospheric resultant
- 3852 N
- Resultant per fastener (equal share)
- 321 N
- Peak bending stress
- 7.599 MPa
- Screening allowable stress
- 51.25 MPa
- Center deflection
- 8.114e-3 mm
- Material safety factor
- 26.98
- Required safety factor
- 4
- O-ring squeeze
- CF metal seal
- O-ring gland fill
- CF metal seal
Reading this example
Use the clamped edge result only if the joint provides the boundary condition. The per-fastener force is a share of the resultant.
Model & assumptions
Atmospheric resultant
F = ΔP πDseal²/4
The atmosphere usually seats an evacuated assembly. Equal force per fastener is a load share only. Determine bolt preload, clamp and flange ratings from the assembly design.
Circular plate
σsimple = 3(3+ν)ΔPr²/(8t²) · σclamped = 3ΔPr²/(4t²)
The simple-support maximum is at the center. For an ideal clamp, maximum bending is at the edge. Deflection follows small-deflection thin-plate theory. The screen uses safety factor 4 for steel and 5 for brittle viewports.
Elastomer gland
Squeeze = 1 − h/d · Fill = (πd²/4)/(hw)
Use assembled clearance h for axial or radial compression. The screen accepts 15% to 30% squeeze and fill below 85%. Fill at or above 100% exceeds the modeled gland capacity. CF copper gasket checks use assembly data.