Custom ADC Alignment of +2 dBu at −16 dBFS
A test tone measures +2 dBu at the converter input while the digital reading is −16 dBFS. Enter that alignment and test an incoming +20 dBu sine.
- Signal level
- 20 dBu
- Input scale
- dBu
- Alignment
- Custom analog and digital reference
- Analog reference level
- 2 dB
- Analog reference unit
- dBu
- Digital reference level
- -16 dBFS
Results
- Mapped digital level
- 2dBFS
- Analog voltage for 0 dBFS
- 6.156Vrms
- Remaining digital headroom
- -2dB
- RMS voltage
- 7.75Vrms
- Level referenced to 0.775 Vrms
- 20dBu
- Level referenced to 1 Vrms
- 17.79dBV
- Sine peak-to-peak voltage
- 21.92031Vpp
- Analog reference voltage
- 0.9757Vrms
- Calibration limit
- Above full scale
The dashed marker indicates calibration. Illuminated segments track the amplitude on a −60 to 0 dBFS display.
Calibration landmarks
| Digital level (dBFS) | Voltage (Vrms) | Analog level (dBu) |
|---|---|---|
| 0 | 6.15604 | 18 |
| -6 | 3.08533 | 12 |
| -12 | 1.54633 | 6 |
| -16 | 0.97567 | 2 |
| -24 | 0.38842 | -6 |
| -48 | 0.02451 | -30 |
Worked calculation
Convert the reference tone to voltage
Vref = 0.775 × 10^(2 ÷ 20) = 0.975667 Vrms
Find the analog voltage at digital full scale
V0dBFS = 0.975667 ÷ 10^(-16 ÷ 20) = 6.15604 Vrms
Map the voltage to the selected digital reference
Digital level = 20 log₁₀(7.75 ÷ 6.156044) = 2 dBFS
What the example shows
The implied full-scale analog level is +18 dBu. A +20 dBu input maps 2 dB above that limit, which the meter marks as over scale.
Model assumptions
- Peak-to-peak conversion assumes a clean sine wave. Program material has a changing crest factor and dBFS calibration here does not measure LUFS or true peak.
- The EBU and SMPTE presets pair their digital reference levels with common analog alignment voltages. Use Custom to enter the reference measured on your converter.
- An analog signal can map above 0 dBFS. This flags the amount above the chosen converter calibration limit.
Common questions
Why does the +20 dBu example show a positive dBFS mapping?
The analog sine is above the voltage implied by 0 dBFS in this setup. A digital converter at that fixed gain cannot represent the full amplitude cleanly.
Can I use an analog measurement at −16 dBFS to infer the full-scale voltage?
Yes, for a converter with linear gain before clipping. Multiply the measured voltage by the amplitude ratio for 16 dB.