Scale a 60-Minute Run by Rotor k Factor
Translate a 60-minute reference run at k = 200 to a different rotor geometry and actual speed.
RCF = 1.118e−5 × r_cm × RPM²
Bench result
Calculated in this browser
- Calculate
- RCF from speed
- Reference radius
- r_max (tube bottom)
- Speed (rpm)
- 30000
- Radius unit
- mm
- Maximum radius
- 100
- Minimum radius
- 50
- Reference rotor k factor
- 200
- Reference time (min)
- 60
- RCF at r_max
- 100,620× g
- Speed
- 30,000rpm
- Rotor k factor
- 195.085
- Scaled run time
- 58.5254min
| Position | Radiusmm | RCF× g |
|---|---|---|
| r_min | 50 | 50,310 |
| r_avg | 75 | 75,465 |
| r_max | 100 | 100,620 |
Worked calculation
Convert reference radius to centimeters
r_max = 100 mm = 10 cm.
Square the speed
RCF = 1.118e−5 × 10 × 30,000² = 100,620 × g.
Estimate the k factor
k = 10¹³ × ln(100 ÷ 50) ÷ (3600 × ω²) = 195.085 at ω = 3,141.59 rad/s.
Scale the reference run
t₂ = 60 × 195.085 ÷ 200 = 58.5254 min.
- Check the instrument, rotor and tube speed ratings. The k-factor estimate uses the entered geometry and speed. Compare runs at the same sedimentation conditions.
How the model works
Enter the distance from the rotation axis to the relevant point in the tube. Use the same radius definition as the protocol or instrument specification.
Frequently asked questions
Which k factor belongs in the numerator for the new run?
Use the new rotor’s k at its actual speed. New time equals reference time multiplied by new k divided by reference k.
Can I scale the run if the sample viscosity also changes?
The simple k ratio assumes the same sedimentation coefficient. A solvent change can affect the coefficient and alter the time estimate.