Scale a 60-Minute Run by Rotor k Factor

Translate a 60-minute reference run at k = 200 to a different rotor geometry and actual speed.

RCF = 1.118e−5 × r_cm × RPM²

Inputs

Optional for speed or RCF alone
Blank uses the midpoint
Optional
Optional with reference time
Optional with reference k factor

Bench result

Calculated in this browser

RCF at r_max
100,620× g
Speed
30,000rpm
Rotor k factor
195.085
Scaled run time
58.5254min
RCF across rotor radius50,30062,90075,50088,000101,0005062.57587.5100RCF (× g)Rotor radius (mm)r_max (100, 100,620)RCF at 30,000 rpm
RCF across rotor radius
RCF by radial position
PositionRadiusmmRCF× g
r_min5050,310
r_avg7575,465
r_max100100,620

How the model works

Enter the distance from the rotation axis to the relevant point in the tube. Use the same radius definition as the protocol or instrument specification.

Frequently asked questions

Which k factor belongs in the numerator for the new run?

Use the new rotor’s k at its actual speed. New time equals reference time multiplied by new k divided by reference k.

Can I scale the run if the sample viscosity also changes?

The simple k ratio assumes the same sedimentation coefficient. A solvent change can affect the coefficient and alter the time estimate.