ASME V Article 8 · ET model
Inconel 718 target deeper than 3δ
A 180 kHz setup is assessed for a target 5 mm below the surface of Inconel 718.
Reading the result. At roughly 3.8 skin depths, the modeled current density is below 5%. Review frequency and probe setup.
Setup
Material
% IACS
Use 1 for a nonmagnetic material.
Test conditions
kHz
mm
×
Use 1 for standard or 3 for effective depth.
Results update as you edit. Use calibrated measurements.
Inspection readout
SI units- Skin depth δ
- 1.3165mm
- 3δ depth
- 3.9494mm
- J / J₀
- 2.2414%
- Phase lag
- 217.61°
Beyond 3δ
The idealized current density is below 5% of its surface value.
- Depth / skin depth
- 3.7981×
- Unwrapped phase lag
- 3.7981rad
- Conductivity used
- 0.812MS/m
- Relative conductivity used
- 1.4% IACS
- Frequency for target d/δ
- 112.3kHz
Conductivity presets are nominal at room temperature. Real coil response also depends on lift-off, geometry, frequency response and local permeability. Equations and references