ASME V Article 8 · ET model

Ti-6Al-4V near-surface screening

A 25 kHz coil is modeled on a 1% IACS titanium alloy with a 1 mm target depth.

Reading the result. The low nominal conductivity gives about 4.18 mm of standard penetration in the half-space model.

Setup

Material

Use 1 for a nonmagnetic material.

Test conditions

kHz
mm
×
Use 1 for standard or 3 for effective depth.
Worked examples

Results update as you edit. Use calibrated measurements.

Inspection readout

SI units
Skin depth δ
4.1796mm
3δ depth
12.539mm
J / J₀
78.721%
Phase lag
13.708°
Within 3δ model depth

Skin depth describes current density decay in a homogeneous half-space.

Current decay and impedance phaseThe left plot follows J/J₀ with depth. The right plot wraps phase by 360° and uses the same relative current magnitude.Induction profileδ 4.1796 mmTarget depth / δ0.23926×EC decay3δ depthTarget phase
The left plot follows J/J₀ with depth. The right plot wraps phase by 360° and uses the same relative current magnitude.
Depth / skin depth
0.23926×
Unwrapped phase lag
0.23926rad
Conductivity used
0.58MS/m
Relative conductivity used
1% IACS
Frequency for target d/δ
436.73kHz

Conductivity presets are nominal at room temperature. Real coil response also depends on lift-off, geometry, frequency response and local permeability. Equations and references