Unfocused circular transducer

½ in, 5 MHz circular element in steel

An unfocused 12.7 mm steel-contact transducer is used at 5 MHz for a 65 mm target.

Reading the result. The finite-wavelength near field is about 33.76 mm. The target lies beyond the natural focus.

Setup

Active element

mm
MHz

Sound path

m/s
mm
Worked examples

Results update as you edit. Use calibrated measurements.

Inspection readout

SI units
Near field N
33.76mm
Wavelength
1.184mm
−6 dB angle
2.7466°
−20 dB angle
5.7787°
Target at or beyond N

The cone width uses an unfocused circular-piston approximation.

Circular aperture and beam envelopeThe cone outlines are schematic. The −6 dB width uses a far-field cone extrapolated from the element plane.Aperture profileUnfocused pistonTarget depth / N1.9253×Near focus−6 dB−20 dB
The cone outlines are schematic. The −6 dB width uses a far-field cone extrapolated from the element plane.
−6 dB cone diameter at target
6.2367mm
Target depth / near-field length
1.9253×
D²/(4λ) reference
34.056mm

Use an unfocused, circular active element in a homogeneous medium. Focused probes, phased-array apertures and measured beam maps require a different model. Equations and references