Unfocused circular transducer
Low-frequency immersion stand-off
A 10 mm unfocused element runs at 1 MHz in a nominal 1,500 m/s water path with a 30 mm reflector distance.
Reading the result. The wavelength increases to 1.5 mm. In this model the broader cone is evaluated beyond the near field.
Setup
Active element
mm
MHz
Sound path
m/s
mm
Results update as you edit. Use calibrated measurements.
Inspection readout
SI units- Near field N
- 16.292mm
- Wavelength
- 1.5mm
- −6 dB angle
- 4.4219°
- −20 dB angle
- 9.323°
Target at or beyond N
The cone width uses an unfocused circular-piston approximation.
- −6 dB cone diameter at target
- 4.6398mm
- Target depth / near-field length
- 1.8414×
- D²/(4λ) reference
- 16.667mm
Use an unfocused, circular active element in a homogeneous medium. Focused probes, phased-array apertures and measured beam maps require a different model. Equations and references