Titanium fixture temperature correction
JCGM 100 / ASME B89.7.3.1 / SI units
- Combined uᶜ
- 2.270706 µm
- Expanded U
- 4.552603 µm
- Coverage k
- 2.00493
- TUR
- 4.39309
Inputs
- Measured result
- 160.014 mm
- Reference length
- 160 mm
- Lower specification
- 159.98 mm
- Upper specification
- 160.02 mm
- Decision rule
- Guard band · ±U
- Coverage factor
- Student-t · 95.45%
- Repeatability SD
- 2.2 µm
- Repeat count
- 8
- Instrument input
- Expanded certificate uncertainty
- Instrument uncertainty / MPE
- 3.6 µm
- Certificate k
- 2
- Instrument df (optional)
- —
- Resolution increment
- 1 µm
- Probe / form half-width
- 1.4 µm
- Fixture / alignment half-width
- 1.1 µm
- Workpiece temperature
- 24.8 °C
- Estimated differential CTE
- −2.9 ppm/°C
- Differential CTE uncertainty
- 0.9 ppm/°C
- Temperature uncertainty
- 0.4 °C
- Thermal input distributions
- Rectangular half-widths
- Apply nominal thermal correction
- Yes
Uncertainty budget
| Component | Standard u | Degrees of freedom | Variance share |
|---|---|---|---|
| Repeatability | 0.7778175 µm | 7 | 11.734 % |
| Instrument | 1.8 µm | ∞ | 62.838 % |
| Resolution | 0.2886751 µm | ∞ | 1.6162 % |
| Probe / form | 0.8082904 µm | ∞ | 12.671 % |
| Fixture / alignment | 0.6350853 µm | ∞ | 7.8224 % |
| Thermal mismatch | 0.4136467 µm | ∞ | 3.3185 % |
Conformance and coverage
| Check | Result |
|---|---|
| Measured result used | 160.0162272 mm |
| Applied thermal correction | 2.2272 µm |
| Reported interval | 160.0116746 mm to 160.0207798 mm |
| Guarded acceptance zone | 159.9845526 mm to 160.0154474 mm |
| Simple acceptance | Accepted |
| Guarded result | Indeterminate |
| Closest decision margin | −0.7798033 µm |
| Effective degrees of freedom | 508.429 |
| Coverage basis | 95.45 % two-sided |
Decision based on the measured interval, including expanded uncertainty. The illustrated three-zone guard rule accepts values whose full ±U interval lies inside spec, rejects those wholly outside, and flags overlap. A contract may specify another rule, correlations, temperature correction or uncertainty coverage.