Semiconductor fixture gap
JCGM 100 / ASME B89.7.3.1 / SI units
- Combined uᶜ
- 2.811609 µm
- Expanded U
- 5.623219 µm
- Coverage k
- 2
- TUR
- 0.533502
Inputs
- Measured result
- 0.202 mm
- Reference length
- 25 mm
- Lower specification
- 0.198 mm
- Upper specification
- 0.204 mm
- Decision rule
- Guard band · ±U
- Coverage factor
- Fixed k = 2
- Repeatability SD
- 1.6 µm
- Repeat count
- 5
- Instrument input
- Standard uncertainty
- Instrument uncertainty / MPE
- 2.5 µm
- Instrument df (optional)
- —
- Resolution increment
- 1 µm
- Probe / form half-width
- 1.4 µm
- Fixture / alignment half-width
- 1.1 µm
- Workpiece temperature
- 22.5 °C
- Estimated differential CTE
- 2.7 ppm/°C
- Differential CTE uncertainty
- 1.4 ppm/°C
- Temperature uncertainty
- 0.6 °C
- Thermal input distributions
- Rectangular half-widths
- Apply nominal thermal correction
- No
Uncertainty budget
| Component | Standard u | Degrees of freedom | Variance share |
|---|---|---|---|
| Repeatability | 0.7155418 µm | 4 | 6.4768 % |
| Instrument | 2.5 µm | ∞ | 79.062 % |
| Resolution | 0.2886751 µm | ∞ | 1.0542 % |
| Probe / form | 0.8082904 µm | ∞ | 8.2647 % |
| Fixture / alignment | 0.6350853 µm | ∞ | 5.1022 % |
| Thermal mismatch | 0.05610556 µm | ∞ | 0.03982 % |
Conformance and coverage
| Check | Result |
|---|---|
| Measured result used | 0.202 mm |
| Applied thermal correction | 0 µm |
| Reported interval | 0.19637678105 mm to 0.20762321895 mm |
| Guarded acceptance zone | Empty |
| Simple acceptance | Accepted |
| Guarded result | Indeterminate |
| Closest decision margin | −3.623219 µm |
| Effective degrees of freedom | 953.543 |
| Coverage basis | Fixed k = 2 |
Decision based on the measured interval, including expanded uncertainty. The illustrated three-zone guard rule accepts values whose full ±U interval lies inside spec, rejects those wholly outside, and flags overlap. A contract may specify another rule, correlations, temperature correction or uncertainty coverage.