Turbo foreline & backing pump / Worked example
Nitrogen Purge at Idle
Process flow is off while a small nitrogen purge remains active. The backing pump still runs at 9 L/s.
Inputs
Compression and foreline gas
Live readout
Backing pressure screen pass- Total foreline throughput
- 0.018 Pa·m³/s
- Estimated foreline pressure
- 2 Pa
- Pressure limit with headroom
- 45 Pa
- Compression-limited inlet pressure
- 2e-9 Pa
- Backing pressure / maximum
- 3.333%
- Process share of total load
- 0%
- Foreline Knudsen number
- 0.2149
- Foreline flow regime
- Transitional flow
Reading this example
The purge sets the idle pressure even with zero process load. The process share readout should be zero.
Model & assumptions
Foreline balance
Pfore = (Qproc + Qpurge)/Sback,eff
The reserve screen sets a lower operating limit than the maximum backing pressure. Input the effective speed at the foreline port and this pressure range.
Compression floor
Pin,min = Pfore/K₀
Compression is species-specific and depends on operating conditions. This screen applies the entered foreline pressure to the selected species as a worst-case partial pressure. Use measured partial pressure where known.