Turbo foreline & backing pump / Worked example
Turbo Foreline Stall Risk
A small backing pump sees a 0.49 Pa·m³/s transient process and purge load. Its nominal effective speed is 3.5 L/s.
Inputs
Compression and foreline gas
Live readout
Foreline stall risk- Total foreline throughput
- 0.49 Pa·m³/s
- Estimated foreline pressure
- 140 Pa
- Pressure limit with headroom
- 80 Pa
- Compression-limited inlet pressure
- 1.4e-7 Pa
- Backing pressure / maximum
- 140%
- Process share of total load
- 91.84%
- Foreline Knudsen number
- 1.964e-3
- Foreline flow regime
- Viscous / continuum
Reading this example
The modeled pressure crosses the entered 100 Pa maximum. Compare actual process duty and the backing pump curve at that pressure.
Model & assumptions
Foreline balance
Pfore = (Qproc + Qpurge)/Sback,eff
The reserve screen sets a lower operating limit than the maximum backing pressure. Input the effective speed at the foreline port and this pressure range.
Compression floor
Pin,min = Pfore/K₀
Compression is species-specific and depends on operating conditions. This screen applies the entered foreline pressure to the selected species as a worst-case partial pressure. Use measured partial pressure where known.