Turbo Foreline Pressure, Backing Pump & Compression Ratio

Estimate turbopump backing pressure, headroom, light-gas compression floor and the Knudsen regime in a foreline.

Readout ↓

Inputs

Backing load and pressure limit
Pa·m³/s
Pa·m³/s
L/s
Pa
%
Compression and foreline gas
mm
°C

Live readout

Backing pressure screen pass
Foreline P69.33PaReserve limit75PaInlet floor6.933e-3PaBacking load69.33%PᵢₙChamberK₀TurboPꜰForelineSBackingLog pressure cascadeInlet floor6.93e-3 PaForeline69.3 PaMax backing100 PaReserve limit75 PaBacking load69.3%Compression1e4
Total foreline throughput
0.208 Pa·m³/s
Estimated foreline pressure
69.33 Pa
Pressure limit with headroom
75 Pa
Compression-limited inlet pressure
6.933e-3 Pa
Backing pressure / maximum
69.33%
Process share of total load
86.54%
Foreline Knudsen number
3.967e-3
Foreline flow regime
Viscous / continuum

    Model & assumptions

    Foreline balance

    Pfore = (Qproc + Qpurge)/Sback,eff

    The reserve screen sets a lower operating limit than the maximum backing pressure. Input the effective speed at the foreline port and this pressure range.

    Compression floor

    Pin,min = Pfore/K₀

    Compression is species-specific and depends on operating conditions. This screen applies the entered foreline pressure to the selected species as a worst-case partial pressure. Use measured partial pressure where known.