Helium Leak Rate, Detector Response & Rate of Rise

Convert tracer helium to an equivalent gas leak, estimate 1τ/3τ/5τ response and calculate the isolation rate-of-rise gas load.

Readout ↓

Inputs

Tracer and detector
Pa·m³/s
%
L
L/s
Rate-of-rise isolation test
Pa
s

Live readout

Molecular helium trace
Equivalent Q1.487e-9Pa·m³/s1τ response5.067s3τ response15.2sRate-of-rise Q2.217e-7Pa·m³/sHeTracer mix10%VTest volume38 LSDetector7.5 L/sDetector signal · % of final response0%50%100%01τ3τ5τ1τ · 63.2%5.07 s3τ · 95.0%15.2 s5τ · 99.3%25.3 s
100% helium equivalent
4e-9 Pa·m³/s
Gas equivalence multiplier
0.3717
Equivalent gas throughput
1.487e-9 Pa·m³/s
63.2% response (1τ)
5.067 s
95.0% response (3τ)
15.2 s
99.3% response (5τ)
25.33 s
Isolation gas load
2.217e-7 Pa·m³/s

    Model & assumptions

    Tracer correction

    Qgas = (QHe/fHe) × √(MHe/Mgas) (molecular)

    For a laminar viscous leak path use ηHe/ηgas with representative viscosities at 20 °C. These ratios assume identical geometry, pressure conditions and temperature.

    Detector and isolation

    τ = V/SHe · Signal = 1 − e⁻ᵗ/ᵗᵃᵘ · Qrise = V ΔP/Δt

    The signal curve assumes a well-mixed volume and first-order response. An isolated chamber can rise from desorption or permeation as well as an external leak.