Helium leak & response time / Worked example

Large Chamber Rate-of-Rise Test

A 900 L chamber is isolated for 180 s and rises by 0.008 Pa. Compare the inferred total load with its helium trace.

Readout ↓

Inputs

Tracer and detector
Pa·m³/s
%
L
L/s
Rate-of-rise isolation test
Pa
s

Live readout

Molecular helium trace
Equivalent Q2.23e-10Pa·m³/s1τ response75s3τ response225sRate-of-rise Q4e-5Pa·m³/sHeTracer mix100%VTest volume900 LSDetector12 L/sDetector signal · % of final response0%50%100%01τ3τ5τ1τ · 63.2%75 s3τ · 95.0%225 s5τ · 99.3%375 s
100% helium equivalent
6e-10 Pa·m³/s
Gas equivalence multiplier
0.3717
Equivalent gas throughput
2.23e-10 Pa·m³/s
63.2% response (1τ)
75 s
95.0% response (3τ)
225 s
99.3% response (5τ)
375 s
Isolation gas load
4e-5 Pa·m³/s

    Reading this example

    A pressure rise may come from desorption as well as a leak. Compare repeat runs after conditioning and use a tracer to localize a defect.

    Model & assumptions

    Tracer correction

    Qgas = (QHe/fHe) × √(MHe/Mgas) (molecular)

    For a laminar viscous leak path use ηHe/ηgas with representative viscosities at 20 °C. These ratios assume identical geometry, pressure conditions and temperature.

    Detector and isolation

    τ = V/SHe · Signal = 1 − e⁻ᵗ/ᵗᵃᵘ · Qrise = V ΔP/Δt

    The signal curve assumes a well-mixed volume and first-order response. An isolated chamber can rise from desorption or permeation as well as an external leak.