Helium leak & response time / Worked example
Small-Volume Detector Dwell
A 0.42 L manifold sees 2.8 L/s effective helium speed. Estimate ideal signal dwell for a short tracer pulse.
Inputs
Rate-of-rise isolation test
Live readout
Molecular helium trace- 100% helium equivalent
- 1.75e-11 Pa·m³/s
- Gas equivalence multiplier
- 0.3717
- Equivalent gas throughput
- 6.505e-12 Pa·m³/s
- 63.2% response (1τ)
- 0.15 s
- 95.0% response (3τ)
- 0.45 s
- 99.3% response (5τ)
- 0.75 s
- Isolation gas load
- 5.6e-9 Pa·m³/s
Reading this example
Fast chamber response can be shorter than detector electronics or transport delay. Verify dwell at the actual test port.
Model & assumptions
Tracer correction
Qgas = (QHe/fHe) × √(MHe/Mgas) (molecular)
For a laminar viscous leak path use ηHe/ηgas with representative viscosities at 20 °C. These ratios assume identical geometry, pressure conditions and temperature.
Detector and isolation
τ = V/SHe · Signal = 1 − e⁻ᵗ/ᵗᵃᵘ · Qrise = V ΔP/Δt
The signal curve assumes a well-mixed volume and first-order response. An isolated chamber can rise from desorption or permeation as well as an external leak.