Helium leak & response time / Worked example

Viscous Nitrogen-Equivalent Leak

A coarse leak path is modeled as laminar viscous. Compare full helium tracer flow with nitrogen at identical pressure conditions.

Readout ↓

Inputs

Tracer and detector
Pa·m³/s
%
L
L/s
Rate-of-rise isolation test
Pa
s

Live readout

Viscous helium trace
Equivalent Q2.784e-7Pa·m³/s1τ response5.294s3τ response15.88sRate-of-rise Q1.05e-7Pa·m³/sHeTracer mix100%VTest volume18 LSDetector3.4 L/sDetector signal · % of final response0%50%100%01τ3τ5τ1τ · 63.2%5.29 s3τ · 95.0%15.9 s5τ · 99.3%26.5 s
100% helium equivalent
2.5e-7 Pa·m³/s
Gas equivalence multiplier
1.114
Equivalent gas throughput
2.784e-7 Pa·m³/s
63.2% response (1τ)
5.294 s
95.0% response (3τ)
15.88 s
99.3% response (5τ)
26.47 s
Isolation gas load
1.05e-7 Pa·m³/s

    Reading this example

    Viscosity ratios apply to this assumed flow regime. A small high-vacuum defect may require molecular equivalence instead.

    Model & assumptions

    Tracer correction

    Qgas = (QHe/fHe) × √(MHe/Mgas) (molecular)

    For a laminar viscous leak path use ηHe/ηgas with representative viscosities at 20 °C. These ratios assume identical geometry, pressure conditions and temperature.

    Detector and isolation

    τ = V/SHe · Signal = 1 − e⁻ᵗ/ᵗᵃᵘ · Qrise = V ΔP/Δt

    The signal curve assumes a well-mixed volume and first-order response. An isolated chamber can rise from desorption or permeation as well as an external leak.