Mean free path & Knudsen number / Worked example
UHV Wafer Transfer Chamber
A 160 mm opening joins two wafer transfer modules at 3 × 10⁻⁸ Pa. Estimate nitrogen transport and the time for a clean surface to accumulate a monolayer.
Inputs
Surface coverage
Live readout
Molecular flow- Mean free path λ
- 2.292e8 mm
- Knudsen number Kn
- 1.432e6
- Gas number density
- 7.412e12 m⁻³
- Mean thermal speed
- 470.7 m/s
- Wall impingement flux
- 8.722e14 m⁻² s⁻¹
- Monolayer time
- 5.732e4 s
- Molecular below
- 0.08594 Pa
- Continuum above
- 4.297 Pa
Reading this example
Use the impingement flux to compare exposure at the transfer pressure. A sticking coefficient below one increases the deposition time.
Model & assumptions
Mean free path and regime
λ = kBT / (√2 πd²P) · Kn = λ/D
At Kn below 0.01 use continuum screening. Kn from 0.01 through 0.5 is transitional. Above 0.5, use molecular screening. The transition pressures move with gas temperature, species and channel size.
Wall flux and coverage
n = P/(kBT) · v̄ = √(8RT/πM) · J = nv̄/4 · τ = Nₛ/(sJ)
The Maxwell–Boltzmann mean speed assumes a thermalized ideal gas. The wall flux counts molecular hits. Multiply by the sticking probability to estimate adsorption on an initially clean surface.