Gas Kinetics, Mean Free Path & Knudsen Flow Regime
Find the Knudsen number, mean free path, impingement flux and monolayer formation time for gas in a vacuum chamber or channel.
Inputs
Surface coverage
Live readout
Molecular flow- Mean free path λ
- 114.6 mm
- Knudsen number Kn
- 2.292
- Gas number density
- 1.482e19 m⁻³
- Mean thermal speed
- 470.7 m/s
- Wall impingement flux
- 1.744e21 m⁻² s⁻¹
- Monolayer time
- 5.732e-3 s
- Molecular below
- 0.275 Pa
- Continuum above
- 13.75 Pa
Model & assumptions
Mean free path and regime
λ = kBT / (√2 πd²P) · Kn = λ/D
At Kn below 0.01 use continuum screening. Kn from 0.01 through 0.5 is transitional. Above 0.5, use molecular screening. The transition pressures move with gas temperature, species and channel size.
Wall flux and coverage
n = P/(kBT) · v̄ = √(8RT/πM) · J = nv̄/4 · τ = Nₛ/(sJ)
The Maxwell–Boltzmann mean speed assumes a thermalized ideal gas. The wall flux counts molecular hits. Multiply by the sticking probability to estimate adsorption on an initially clean surface.