Mean free path & Knudsen number / Worked example

Water Vapor in a Load Lock

A recently opened load lock is pumping through 0.09 Pa. Estimate the water vapor impingement rate at a 95 mm characteristic dimension.

Readout ↓

Inputs

Gas and geometry
Pa
°C
mm
Surface coverage
Enter a probability from 0 to 1.
m⁻²

Live readout

Molecular flow
Mean free path149.1mmKnudsen no.1.569Thermal speed596.9m/sMonolayer time5.194e-3sMolecule pathsD 95 mmLog pressure ladder · Torr1e31e01e-41e-81e-11Continuum >14.1 PaOperating P0.09 PaMolecular <0.282 Pa
Mean free path λ
149.1 mm
Knudsen number Kn
1.569
Gas number density
2.15e19 m⁻³
Mean thermal speed
596.9 m/s
Wall impingement flux
3.209e21 m⁻² s⁻¹
Monolayer time
5.194e-3 s
Molecular below
0.2824 Pa
Continuum above
14.12 Pa

    Reading this example

    A high wall flux signals frequent impacts. Coverage still depends on temperature, adsorption chemistry and the actual partial pressure.

    Model & assumptions

    Mean free path and regime

    λ = kBT / (√2 πd²P) · Kn = λ/D

    At Kn below 0.01 use continuum screening. Kn from 0.01 through 0.5 is transitional. Above 0.5, use molecular screening. The transition pressures move with gas temperature, species and channel size.

    Wall flux and coverage

    n = P/(kBT) · v̄ = √(8RT/πM) · J = nv̄/4 · τ = Nₛ/(sJ)

    The Maxwell–Boltzmann mean speed assumes a thermalized ideal gas. The wall flux counts molecular hits. Multiply by the sticking probability to estimate adsorption on an initially clean surface.