Pipe conductance & pump speed / Worked example

Thin Aperture at an RGA Port

A 10 mm circular aperture limits the gas feed toward a residual gas analyzer at 10⁻⁴ Pa.

Readout ↓

Inputs

Tube and pump
mm
m
Use zero for a thin aperture in molecular flow.
L/s
Pa
Gas and series restriction
°C
L/s
Enter 0 to omit a valve, baffle or other restriction.

Live readout

Molecular flow
Tube C9.089L/sChamber speed3.614L/sUtilization60.24%ID for 80%16.25mmChamberTubePumpID 10 mmChamber speed · L/s vs ID · mm03609.118.2FlowMolecularPump speed6 L/s80% target4.8 L/s
Molecular conductance
9.089 L/s
Laminar conductance model
Thin aperture
Molecular transmission
100%
Knudsen number Kn
6838
Regime-adjusted tube conductance
9.089 L/s
Total series conductance
9.089 L/s
Effective chamber speed
3.614 L/s
Pump utilization
60.24%
Minimum inside diameter for 80%
16.25 mm

    Reading this example

    At zero modeled thickness the transmission correction becomes one. The aperture flux alone sets its molecular conductance.

    Model & assumptions

    Tube throughput

    Cₘ = (A v̄/4) / (1 + 3L/4D) · Cᵥ = πD⁴P̄/(128ηL)

    The aperture-to-long-tube resistance approximation approaches Clausing’s long-tube limit. It is a simplified short-tube correction. The viscous model assumes isothermal, fully developed laminar flow.

    Chamber speed

    1/Cₜ = 1/C + 1/C₂ · 1/Sₑ = 1/Sₚ + 1/Cₜ

    Transition uses a logarithmic Knudsen weight between the two limits. The 80% search requires Cₜ ≥ 4Sₚ. A fixed series restriction at or below 4Sₚ makes the 80% target unattainable by enlarging the tube alone.