Pipe conductance & pump speed / Worked example

Transitional Deposition Exhaust

An argon deposition line runs at mean pressure 0.8 Pa with a 32 mm diameter and 18 cm straight run.

Readout ↓

Inputs

Tube and pump
mm
m
Use zero for a thin aperture in molecular flow.
L/s
Pa
Gas and series restriction
°C
L/s
Enter 0 to omit a valve, baffle or other restriction.

Live readout

Transitional flow
Tube C13.94L/sChamber speed12.16L/sUtilization12.8%ID for 80%98.82mmChamberTubePumpID 32 mmChamber speed · L/s vs ID · mm047.595055.3111FlowTransitionPump speed95 L/s80% target76 L/s
Molecular conductance
15.19 L/s
Laminar conductance model
5.129 L/s
Molecular transmission
19.16%
Knudsen number Kn
0.3078
Regime-adjusted tube conductance
13.94 L/s
Total series conductance
13.94 L/s
Effective chamber speed
12.16 L/s
Pump utilization
12.8%
Minimum inside diameter for 80%
98.82 mm

    Reading this example

    Compare the two limiting conductances before relying on the transition blend. A measured speed curve gives better process accuracy.

    Model & assumptions

    Tube throughput

    Cₘ = (A v̄/4) / (1 + 3L/4D) · Cᵥ = πD⁴P̄/(128ηL)

    The aperture-to-long-tube resistance approximation approaches Clausing’s long-tube limit. It is a simplified short-tube correction. The viscous model assumes isothermal, fully developed laminar flow.

    Chamber speed

    1/Cₜ = 1/C + 1/C₂ · 1/Sₑ = 1/Sₚ + 1/Cₜ

    Transition uses a logarithmic Knudsen weight between the two limits. The 80% search requires Cₜ ≥ 4Sₚ. A fixed series restriction at or below 4Sₚ makes the 80% target unattainable by enlarging the tube alone.