Pipe conductance & pump speed / Worked example
Thin Aperture at an RGA Port
A 10 mm circular aperture limits the gas feed toward a residual gas analyzer at 10⁻⁴ Pa.
Inputs
Gas and series restriction
Live readout
Molecular flow- Molecular conductance
- 9.089 L/s
- Laminar conductance model
- Thin aperture
- Molecular transmission
- 100%
- Knudsen number Kn
- 6838
- Regime-adjusted tube conductance
- 9.089 L/s
- Total series conductance
- 9.089 L/s
- Effective chamber speed
- 3.614 L/s
- Pump utilization
- 60.24%
- Minimum inside diameter for 80%
- 16.25 mm
Reading this example
At zero modeled thickness the transmission correction becomes one. The aperture flux alone sets its molecular conductance.
Model & assumptions
Tube throughput
Cₘ = (A v̄/4) / (1 + 3L/4D) · Cᵥ = πD⁴P̄/(128ηL)
The aperture-to-long-tube resistance approximation approaches Clausing’s long-tube limit. It is a simplified short-tube correction. The viscous model assumes isothermal, fully developed laminar flow.
Chamber speed
1/Cₜ = 1/C + 1/C₂ · 1/Sₑ = 1/Sₚ + 1/Cₜ
Transition uses a logarithmic Knudsen weight between the two limits. The 80% search requires Cₜ ≥ 4Sₚ. A fixed series restriction at or below 4Sₚ makes the 80% target unattainable by enlarging the tube alone.