Flange load & viewport stress / Worked example

Axial Gland Fill Above the Screen

A fused silica carrier has an adequate plate but a narrow axial O-ring gland, giving high fill below the physical capacity limit.

Readout ↓

Inputs

Plate and atmospheric load
Pa
Pa
mm
mm
mm
MPa
Enter 0 for a clamp or to omit equal load sharing.
Seal and gland
mm
mm
mm

Live readout

O-ring gland screen fail
Net force877.4NBending stress3.01MPaDeflection2.331e-3mmSafety factor15.95Seal Ø 105 mmViewport safety factorSeal squeeze / gland fillActual SF15.9Minimum SF5Gland fill87.9%
Pressure differential
1.013e5 Pa
Atmospheric resultant
877.4 N
Resultant per fastener (equal share)
146.2 N
Peak bending stress
3.01 MPa
Screening allowable stress
9.6 MPa
Center deflection
2.331e-3 mm
Material safety factor
15.95
Required safety factor
5
O-ring squeeze
22.1%
O-ring gland fill
87.87%
  • Check assembled gland geometry against 15–30% squeeze and below 85% fill.

Reading this example

Squeeze can pass while fill fails. Increasing gland width changes volume without changing nominal squeeze.

Model & assumptions

Atmospheric resultant

F = ΔP πDseal²/4

The atmosphere usually seats an evacuated assembly. Equal force per fastener is a load share only. Determine bolt preload, clamp and flange ratings from the assembly design.

Circular plate

σsimple = 3(3+ν)ΔPr²/(8t²) · σclamped = 3ΔPr²/(4t²)

The simple-support maximum is at the center. For an ideal clamp, maximum bending is at the edge. Deflection follows small-deflection thin-plate theory. The screen uses safety factor 4 for steel and 5 for brittle viewports.

Elastomer gland

Squeeze = 1 − h/d · Fill = (πd²/4)/(hw)

Use assembled clearance h for axial or radial compression. The screen accepts 15% to 30% squeeze and fill below 85%. Fill at or above 100% exceeds the modeled gland capacity. CF copper gasket checks use assembly data.