Flange load & viewport stress / Worked example
ISO-KF Borosilicate Viewport and Gland
A small borosilicate window uses a 3.53 mm O-ring in a 2.8 mm high axial gland.
Inputs
Seal and gland
Live readout
Viewport and seal screen pass- Pressure differential
- 1.013e5 Pa
- Atmospheric resultant
- 267.7 N
- Resultant per fastener (equal share)
- 66.93 N
- Peak bending stress
- 1.9 MPa
- Screening allowable stress
- 5 MPa
- Center deflection
- 8.015e-4 mm
- Material safety factor
- 13.16
- Required safety factor
- 5
- O-ring squeeze
- 20.68%
- O-ring gland fill
- 75.98%
Reading this example
Check atmospheric stress and gland volume together. A nominally adequate squeeze can still leave too little free volume.
Model & assumptions
Atmospheric resultant
F = ΔP πDseal²/4
The atmosphere usually seats an evacuated assembly. Equal force per fastener is a load share only. Determine bolt preload, clamp and flange ratings from the assembly design.
Circular plate
σsimple = 3(3+ν)ΔPr²/(8t²) · σclamped = 3ΔPr²/(4t²)
The simple-support maximum is at the center. For an ideal clamp, maximum bending is at the edge. Deflection follows small-deflection thin-plate theory. The screen uses safety factor 4 for steel and 5 for brittle viewports.
Elastomer gland
Squeeze = 1 − h/d · Fill = (πd²/4)/(hw)
Use assembled clearance h for axial or radial compression. The screen accepts 15% to 30% squeeze and fill below 85%. Fill at or above 100% exceeds the modeled gland capacity. CF copper gasket checks use assembly data.