WeibullMTBFLab · Engineering calculation
Infusion pump bench-service downtime
A hypothetical bench pool models 24,000 hours of uptime between drive repairs, with ten repair hours and two logistics hours.
Operational availability falls below SLA
Inputs
- Mean uptime MTBF
- 24,000 h
- Corrective repair MTTR
- 10 h
- Additional downtime MDT
- 2 h
- SLA target
- 99.98 %
- Mission time
- 8 h
Results
- Inherent Aᵢ
- 99.95835069 %
- Operational Aₒ
- 99.95002499 %
- Downtime/yr
- 4.377811 h/yr
- Transient A(t)
- 99.97705869 %
- Annual downtime
- 262.6687 min/yr
- Annual SLA budget
- 1.752 h/yr
- SLA headroom
- -2.625811 h/yr
- Operational SLA tier
- 3 nines
Aᵢ = MTBF / (MTBF + MTTR)Corrective repair alone gives 99.95835069%.Aₒ = MTBF / (MTBF + MTTR + MDT)Additional downtime brings operational availability to 99.95002499%.Downtime/yr = 8,760 (1 − Aₒ)Annual downtime is 4.377811 hours or 262.6687 minutes.A(t) = Aᵢ + (1−Aᵢ) exp[−(1/MTBF + 1/MTTR)t]Starting fully up, transient availability at 8 h is 99.97705869%.
MTBF is mean uptime, excluding downtime. Enter only additional preventive and logistics downtime per uptime cycle in MDT. The two-state Markov model uses exponential up and repair durations.