Use the unobstructed opening. ANSI/ASSP Z9.5 practice uses 80–100 FPM as a starting range where appropriate. Select a site-specific target after containment testing.
VAV and bypass operation
Q_reduced = max(W × H_reduced × V_face, Q_min)
VAV uses the entered minimum whenever the sash demand falls below it and during all unoccupied hours. The CAV/bypass comparison keeps working flow throughout. The same unoccupied minimum applies to both sash scenarios.
Q_mean is the occupied/unoccupied average in m³/s. Enter degree-hours as the sum of temperature differences over hours, not HDD/CDD in degree-days. The result is purchased kWh, with an MMBtu energy-equivalent display. The model assumes similar climate-hour distribution across shifts and excludes latent loads, fans and heat recovery.
CleanroomVentLab · Calculation sheet · US customary
Constant-volume bypass hood
Inputs
Clear sash width
5 ft
Working height
18 in
Reduced height
12 in
Target face velocity
90 FPM
Exhaust control
Constant volume / bypass
Occupied hours per year
2,600 h
Heating degree-hours
90,000 °F·h
Cooling degree-hours
30,000 °F·h
Heating efficiency
0.9
Cooling COP
3.2
Heating input energy rate
0.09 /kWh
Cooling input energy rate
0.18 /kWh
Cost display currency
USD · US dollar
Results
Working flow
675 CFM
Reduced flow
675 CFM
Energy saved
0 MMBtu
Cost saved
0 USD
Working face velocity
90 FPM
Reduced face velocity
135 FPM
Unoccupied exhaust
675 CFM
Annual mean flow, working
675 CFM
Annual mean flow, reduced
675 CFM
Working heating input
72.9006 MMBtu
Working cooling input
6.83443 MMBtu
Working total energy
79.735 MMBtu
Reduced total energy
79.735 MMBtu
Working annual cost
2,283.39 USD
Reduced annual cost
2,283.39 USD
Constant volume holds exhaust at the same flow in both sash positions.
For preliminary engineering and EHS / Industrial Hygiene review. Design calculations do not replace review by a licensed professional engineer (PE) and a certified industrial hygienist (CIH), commissioning tests, ISO 14644 classification or certification, or assessment against the applicable NFPA 86 edition.