Extended-hours VAV laboratory

A research lab occupies the hood for more than one shift and sets a 25% unoccupied minimum.

Inputs

Hood opening
m
mm
mm
m/s
Control & schedule
%
Percent of working exhaust. Also applies as the occupied minimum.
h
Make-up air utilities
°C·h
°C·h
Decimal delivered-to-input ratio.
/kWh
/kWh
Enter both prices in the selected currency. Currency selection changes labels.

Live results

Sash positions and annual cost

Sash positions and annual costWORKING FLOW1,555m³/hREDUCED FLOW995.3m³/hENERGY SAVED8,074kWhCOST SAVED1,227EURSash openingsMake-up air cost / yearWorking500 mmReduced320 mmWorking3,761 EURReduced2,534 EURReduced sash flow vs workingReduced flow995.3 m³/hWork 1,555 m³/h
Sash openings follow the entered heights. The bars compare annual sensible make-up air costs for the two positions.

The sash comparison uses a shared VAV minimum. Verify containment at both operating positions.

Working flow
1,555.2 m³/h
Reduced flow
995.328 m³/h
Energy saved
8,074.18 kWh
Cost saved
1,227.28 EUR

Airflow and face velocity

Working face velocity
0.48 m/s
Reduced face velocity
0.48 m/s
Unoccupied exhaust
388.8 m³/h
Annual mean flow, working
1,214.33 m³/h
Annual mean flow, reduced
818.078 m³/h

Make-up air per year

Working heating input
22,622.6 kWh
Working cooling input
2,120.87 kWh
Working total energy
24,743.5 kWh
Reduced total energy
16,669.3 kWh
Working annual cost
3,761.01 EUR
Reduced annual cost
2,533.73 EUR

Worked calculation at the starting values

More time at the managed sash increases savings in the schedule-weighted model.

  1. Working flow = 1.8 × 0.5 × 0.48 = 0.432 m³/s.
  2. Average flow with managed sash = 0.2272438 m³/s including 0.108 m³/s during unoccupied hours.
  3. Sash savings = 3,761.007 − 2,533.731 = 1,227.276 EUR per year.
Starting result snapshot
Working flow
1,555.2 m³/h
Reduced flow
995.328 m³/h
Energy saved
8,074.18 kWh
Cost saved
1,227.28 EUR

Equations and assumptions

Face velocity and sash flow

Q = W × H × V_face

Use the unobstructed opening. ANSI/ASSP Z9.5 practice uses 80–100 FPM as a starting range where appropriate. Select a site-specific target after containment testing.

VAV and bypass operation

Q_reduced = max(W × H_reduced × V_face, Q_min)

VAV uses the entered minimum whenever the sash demand falls below it and during all unoccupied hours. The CAV/bypass comparison keeps working flow throughout. The same unoccupied minimum applies to both sash scenarios.

Annual sensible make-up energy

E_heat = 1.2072 × Q_mean × H_DH ÷ η · E_cool = 1.2072 × Q_mean × C_DH ÷ COP

Q_mean is the occupied/unoccupied average in m³/s. Enter degree-hours as the sum of temperature differences over hours, not HDD/CDD in degree-days. The result is purchased kWh, with an MMBtu energy-equivalent display. The model assumes similar climate-hour distribution across shifts and excludes latent loads, fans and heat recovery.

References

More worked examples

Open calculator defaults