R = thickness / k • Q = A × (Tambient − Tpayload) / R
Area is measured at the wall mid-plane using interior dimensions plus one wall thickness. This case assumes the cavity follows the payload target. The readout also shows the higher heat leak if the whole cavity approaches the coolant phase temperature.
Coolant consumption
Rate = 3.6 × Q(W) / L(kJ/kg)
Dry ice uses 571 kJ/kg, about 246 BTU/lb, at −78.5 °C. The 0 °C and −21 °C phase-change pack enthalpies are adjustable starting values.
Reserve and qualification
treserve = m / rate × (1 − reserve fraction)
Budget for seams, bridges, air exchange, payload preconditioning, and ambient extremes. Verify the actual pack-out by an instrumented test against its approved qualification protocol and payload limits.
What happens if a dry-ice cavity runs much colder than the payload?
The temperature driving heat through the walls increases. Compare the readout for a cavity at the coolant phase temperature and use measured pack-out data for qualification.
How does the reserve affect the predicted hold time?
The reserve sets aside the specified percentage of the coolant’s latent energy. Seams, thermal bridges, venting, and variable temperatures can consume that margin.
ColdChainLab · Engineering calculation
Dry Ice & Insulated Shipper Hold-Time Calculator
Metric SI · https://calcsheetlab.com/coldchainlab/dry-ice-shipper-hold-time/
Inputs
Interior length
0.48 m
Interior width
0.36 m
Interior height
0.32 m
Insulation thickness
65 mm
Insulation preset
Expanded polystyrene (EPS)
Effective conductivity
0.035 W/(m·K)
Ambient temperature
30 °C
Payload target temperature
-20 °C
Latent-energy reserve
25 %
Coolant
Dry ice (−78.5 °C)
Coolant mass
8 kg
Results
Reserve hold
29.21h
Heat leak
32.58W
Coolant use
0.205kg/h
Nominal hold
38.95h
Mid-wall heat-transfer area
1.21m²
Wall R-value
1.86m²·K/W
Latent energy available
4,568kJ
Phase-change enthalpy
571kJ/kg
Coolant phase temperature
-78.5°C
Hold time set aside
9.74h
Ambient-to-payload difference
50K
Heat leak with cavity at coolant phase temp.
70.7W
Dry ice releases CO₂. Use packaging designed to vent it and follow the applicable handling and shipping requirements.
Use these estimates for preliminary engineering. Verify results against validated cold-chain qualification protocols, applicable ASHRAE Standards 15 and 34, and local food and pharmaceutical regulations. Use manufacturer data, measured operating conditions, and approved acceptance criteria before acting on a result.