ASME V Article 7 · ASTM E1444/E709
MT Magnetization Current & Field
Screen a direct or central conductor shot, encircling coil or prod setup. Compare selected amperage and tangential field.
Setup
Magnetization geometry
mm
For coils use the procedure’s effective diameter.
mm
mm
mm
Current and field
A
kA/m
Optional. Overrides the circular estimate.
Results update as you edit. Use calibrated measurements.
Inspection readout
SI units- Min. current
- 600A
- Max. current
- 1600A
- Center target
- 1100A
- Surface field
- 6.8925kA/m
Current within screening band
Verify field direction and strength on the part with the documented technique.
- Nominal coil ampere-turns
- —
- Measured geometry L/D
- —
- L/D used in coil formula
- —
- Field reading basis
- Straight-current estimate
- Straight-current estimate
- 6.8925kA/m
- Typical field comparison
- 2.3873kA/m
- Tangential field check
- At or above 30 Oe
Use the written procedure for the part, equipment and governing edition. Field adequacy is demonstrated on the part. The circular-field estimate uses a straight current at the OD radius. Equations and references
Worked examples
Open a setup to inspect its inputs, then try your own values.