Gas bleed & SCCM conversion / Worked example
Argon Sputter MFC Conversion
A sputter recipe calls for 48 SCCM of argon at a 0 °C reference and 130 L/s effective process speed.
Inputs
Live readout
Flow converted- SCCM
- 48 SCCM
- SLM
- 0.048 SLM
- Torr·L/s
- 0.608 Torr·L/s
- mbar·L/s
- 0.8106 mbar·L/s
- Pa·m³/s
- 0.08106 Pa·m³/s
- mol/s
- 3.569e-5 mol/s
- mg/s
- 1.426 mg/s
- Throughput at chamber temperature
- 0.09887 Pa·m³/s
- Predicted process pressure
- 0.7605 Pa
- Downstream / upstream ratio
- Known flow mode
- Critical pressure ratio
- Known flow mode
- Orifice upstream Kn
- Known flow mode
Reading this example
Compare operating chamber throughput with the standard-flow conversions. The temperature basis changes the conversion to mol/s.
Model & assumptions
Reference flow
ṅ = PₛV̇ₛ/(RTₛ) · Q = ṅRT · ṁ = ṅM
SCCM and SLM use the selected reference at 101,325 Pa. Throughput conversions in the unit table use the reference temperature. Process pressure uses the separately entered chamber temperature.
Orifice choking
r* = (2/(γ+1))^(γ/(γ−1)) · r = P₂/P₁
Isentropic mass flow follows the subsonic pressure-ratio equation until the flow becomes sonic. Below the critical ratio the model holds mass flow at its choked value. Process pressure is chamber throughput divided by effective speed.