Gas bleed & SCCM conversion / Worked example

Helium Flow at a 20 °C Reference

A helium MFC reports 16 SCCM on a 20 °C, 1 atm basis. Convert it to gas throughput and mass flow.

Readout ↓

Inputs

Flow basis and process chamber
°C
L/s

Live readout

Flow converted
Standard flow16SCCMMass flow0.04437mg/sChamber load0.02702Pa·m³/sProcess P1.801PaP₁Gas supplyQMFCP₂Process1.8 PaProcess pressure · Pa vs SCCM01.492.97013.226.4Reference20 °C / 1 atmP₂ / P₁Known flowCritical ratio—
SCCM
16 SCCM
SLM
0.016 SLM
Torr·L/s
0.2027 Torr·L/s
mbar·L/s
0.2702 mbar·L/s
Pa·m³/s
0.02702 Pa·m³/s
mol/s
1.109e-5 mol/s
mg/s
0.04437 mg/s
Throughput at chamber temperature
0.02702 Pa·m³/s
Predicted process pressure
1.801 Pa
Downstream / upstream ratio
Known flow mode
Critical pressure ratio
Known flow mode
Orifice upstream Kn
Known flow mode

    Reading this example

    A different standard temperature gives a different molar flow for the same displayed SCCM. Match the basis on the flow certificate.

    Model & assumptions

    Reference flow

    ṅ = PₛV̇ₛ/(RTₛ) · Q = ṅRT · ṁ = ṅM

    SCCM and SLM use the selected reference at 101,325 Pa. Throughput conversions in the unit table use the reference temperature. Process pressure uses the separately entered chamber temperature.

    Orifice choking

    r* = (2/(γ+1))^(γ/(γ−1)) · r = P₂/P₁

    Isentropic mass flow follows the subsonic pressure-ratio equation until the flow becomes sonic. Below the critical ratio the model holds mass flow at its choked value. Process pressure is chamber throughput divided by effective speed.