SCCM Conversion & Choked Orifice Gas Bleed
Convert SCCM, SLM, vacuum throughput and mass flow. Estimate choked or subsonic orifice flow and process pressure.
Inputs
Live readout
Choked flow- SCCM
- 10.5 SCCM
- SLM
- 0.0105 SLM
- Torr·L/s
- 0.133 Torr·L/s
- mbar·L/s
- 0.1773 mbar·L/s
- Pa·m³/s
- 0.01773 Pa·m³/s
- mol/s
- 7.808e-6 mol/s
- mg/s
- 0.2187 mg/s
- Throughput at chamber temperature
- 0.01903 Pa·m³/s
- Predicted process pressure
- 0.1269 Pa
- Downstream / upstream ratio
- 1.48e-6
- Critical pressure ratio
- 0.5283
- Orifice upstream Kn
- 1.786e-3
Model & assumptions
Reference flow
ṅ = PₛV̇ₛ/(RTₛ) · Q = ṅRT · ṁ = ṅM
SCCM and SLM use the selected reference at 101,325 Pa. Throughput conversions in the unit table use the reference temperature. Process pressure uses the separately entered chamber temperature.
Orifice choking
r* = (2/(γ+1))^(γ/(γ−1)) · r = P₂/P₁
Isentropic mass flow follows the subsonic pressure-ratio equation until the flow becomes sonic. Below the critical ratio the model holds mass flow at its choked value. Process pressure is chamber throughput divided by effective speed.