Turbo foreline & backing pump / Worked example

Narrow Foreline at Low Throughput

A 3 mm gauge branch sees a low foreline throughput while the backing pump runs at 14 L/s.

Readout ↓

Inputs

Backing load and pressure limit
Pa·m³/s
Pa·m³/s
L/s
Pa
%
Compression and foreline gas
mm
°C

Live readout

Backing pressure screen pass
Foreline P1.571e-3PaReserve limit52.5PaInlet floor1.571e-9PaBacking load2.245e-3%PᵢₙChamberK₀TurboPꜰForelineSBackingLog pressure cascadeInlet floor1.57e-9 PaForeline1.57e-3 PaMax backing70 PaReserve limit52.5 PaBacking load2.24e-3%Compression1e6
Total foreline throughput
2.2e-5 Pa·m³/s
Estimated foreline pressure
1.571e-3 Pa
Pressure limit with headroom
52.5 Pa
Compression-limited inlet pressure
1.571e-9 Pa
Backing pressure / maximum
2.245e-3%
Process share of total load
63.64%
Foreline Knudsen number
2859
Foreline flow regime
Molecular flow

    Reading this example

    Check Kn at the branch diameter. Use pressure-appropriate conductance to obtain effective speed from a nominal pump curve.

    Model & assumptions

    Foreline balance

    Pfore = (Qproc + Qpurge)/Sback,eff

    The reserve screen sets a lower operating limit than the maximum backing pressure. Input the effective speed at the foreline port and this pressure range.

    Compression floor

    Pin,min = Pfore/K₀

    Compression is species-specific and depends on operating conditions. This screen applies the entered foreline pressure to the selected species as a worst-case partial pressure. Use measured partial pressure where known.