Pipe conductance & pump speed / Worked example

Helium Transfer Manifold

A 40 mm by 0.65 m tube connects a helium test manifold to a 75 L/s pump at 0.016 Pa.

Readout ↓

Inputs

Tube and pump
mm
m
Use zero for a thin aperture in molecular flow.
L/s
Pa
Gas and series restriction
°C
L/s
Enter 0 to omit a valve, baffle or other restriction.

Live readout

Molecular flow
Tube C29.67L/sChamber speed21.26L/sUtilization28.34%ID for 80%89.11mmChamberTubePumpID 40 mmChamber speed · L/s vs ID · mm037.575049.999.8FlowMolecularPump speed75 L/s80% target60 L/s
Molecular conductance
29.67 L/s
Laminar conductance model
0.07891 L/s
Molecular transmission
7.583%
Knudsen number Kn
21.06
Regime-adjusted tube conductance
29.67 L/s
Total series conductance
29.67 L/s
Effective chamber speed
21.26 L/s
Pump utilization
28.34%
Minimum inside diameter for 80%
89.11 mm

    Reading this example

    The higher mean thermal speed of helium changes its molecular conductance. Enter the pump speed for helium for a consistent comparison.

    Model & assumptions

    Tube throughput

    Cₘ = (A v̄/4) / (1 + 3L/4D) · Cᵥ = πD⁴P̄/(128ηL)

    The aperture-to-long-tube resistance approximation approaches Clausing’s long-tube limit. It is a simplified short-tube correction. The viscous model assumes isothermal, fully developed laminar flow.

    Chamber speed

    1/Cₜ = 1/C + 1/C₂ · 1/Sₑ = 1/Sₚ + 1/Cₜ

    Transition uses a logarithmic Knudsen weight between the two limits. The 80% search requires Cₜ ≥ 4Sₚ. A fixed series restriction at or below 4Sₚ makes the 80% target unattainable by enlarging the tube alone.