Pipe conductance & pump speed / Worked example
Helium Transfer Manifold
A 40 mm by 0.65 m tube connects a helium test manifold to a 75 L/s pump at 0.016 Pa.
Inputs
Gas and series restriction
Live readout
Molecular flow- Molecular conductance
- 29.67 L/s
- Laminar conductance model
- 0.07891 L/s
- Molecular transmission
- 7.583%
- Knudsen number Kn
- 21.06
- Regime-adjusted tube conductance
- 29.67 L/s
- Total series conductance
- 29.67 L/s
- Effective chamber speed
- 21.26 L/s
- Pump utilization
- 28.34%
- Minimum inside diameter for 80%
- 89.11 mm
Reading this example
The higher mean thermal speed of helium changes its molecular conductance. Enter the pump speed for helium for a consistent comparison.
Model & assumptions
Tube throughput
Cₘ = (A v̄/4) / (1 + 3L/4D) · Cᵥ = πD⁴P̄/(128ηL)
The aperture-to-long-tube resistance approximation approaches Clausing’s long-tube limit. It is a simplified short-tube correction. The viscous model assumes isothermal, fully developed laminar flow.
Chamber speed
1/Cₜ = 1/C + 1/C₂ · 1/Sₑ = 1/Sₚ + 1/Cₜ
Transition uses a logarithmic Knudsen weight between the two limits. The 80% search requires Cₜ ≥ 4Sₚ. A fixed series restriction at or below 4Sₚ makes the 80% target unattainable by enlarging the tube alone.