Pipe conductance & pump speed / Worked example

Cryopump Neck With a Baffle

A 180 mm neck, 1.2 m long, connects a high-vacuum chamber to a cryopump through a 5,000 L/s baffle.

Readout ↓

Inputs

Tube and pump
mm
m
Use zero for a thin aperture in molecular flow.
L/s
Pa
Gas and series restriction
°C
L/s
Enter 0 to omit a valve, baffle or other restriction.

Live readout

Molecular flow
Tube C490.8L/sChamber speed292.9L/sUtilization34.46%ID for 80%553.9mmChamberTube + valvePumpID 180 mmChamber speed · L/s vs ID · mm04258500310620FlowMolecularPump speed850 L/s80% target680 L/s
Molecular conductance
490.8 L/s
Laminar conductance model
1.184 L/s
Molecular transmission
16.67%
Knudsen number Kn
37.99
Regime-adjusted tube conductance
490.8 L/s
Total series conductance
446.9 L/s
Effective chamber speed
292.9 L/s
Pump utilization
34.46%
Minimum inside diameter for 80%
553.9 mm

    Reading this example

    The series restriction affects diameter sizing even when its conductance looks large compared with the pump.

    Model & assumptions

    Tube throughput

    Cₘ = (A v̄/4) / (1 + 3L/4D) · Cᵥ = πD⁴P̄/(128ηL)

    The aperture-to-long-tube resistance approximation approaches Clausing’s long-tube limit. It is a simplified short-tube correction. The viscous model assumes isothermal, fully developed laminar flow.

    Chamber speed

    1/Cₜ = 1/C + 1/C₂ · 1/Sₑ = 1/Sₚ + 1/Cₜ

    Transition uses a logarithmic Knudsen weight between the two limits. The 80% search requires Cₜ ≥ 4Sₚ. A fixed series restriction at or below 4Sₚ makes the 80% target unattainable by enlarging the tube alone.