Pipe conductance & pump speed / Worked example
Cryopump Neck With a Baffle
A 180 mm neck, 1.2 m long, connects a high-vacuum chamber to a cryopump through a 5,000 L/s baffle.
Inputs
Gas and series restriction
Live readout
Molecular flow- Molecular conductance
- 490.8 L/s
- Laminar conductance model
- 1.184 L/s
- Molecular transmission
- 16.67%
- Knudsen number Kn
- 37.99
- Regime-adjusted tube conductance
- 490.8 L/s
- Total series conductance
- 446.9 L/s
- Effective chamber speed
- 292.9 L/s
- Pump utilization
- 34.46%
- Minimum inside diameter for 80%
- 553.9 mm
Reading this example
The series restriction affects diameter sizing even when its conductance looks large compared with the pump.
Model & assumptions
Tube throughput
Cₘ = (A v̄/4) / (1 + 3L/4D) · Cᵥ = πD⁴P̄/(128ηL)
The aperture-to-long-tube resistance approximation approaches Clausing’s long-tube limit. It is a simplified short-tube correction. The viscous model assumes isothermal, fully developed laminar flow.
Chamber speed
1/Cₜ = 1/C + 1/C₂ · 1/Sₑ = 1/Sₚ + 1/Cₜ
Transition uses a logarithmic Knudsen weight between the two limits. The 80% search requires Cₜ ≥ 4Sₚ. A fixed series restriction at or below 4Sₚ makes the 80% target unattainable by enlarging the tube alone.