ASME V Article 2 · Ug screen
Co-60 thick section after one half-life
A 114.3 mm reference geometry is repeated at the same 600 mm SFD after one Co-60 half-life.
Reading the result. With the same original activity and geometry, halve the current activity and double the reference exposure time.
Setup
Exposure geometry
mm
mm
Film-side surface in contact.
mm
Reference shot
mm
min
GBq
Activity at the reference shot.
GBq
d
Results update as you edit. Use calibrated measurements.
Inspection readout
SI units- Calculated Ug
- 0.70599mm
- ASME Ug limit
- 1.778mm
- Minimum SFD
- 307.16mm
- Exposure time
- 10min
Pass · within Ug limit
This verdict covers the geometric unsharpness screen only.
- Source-to-object distance
- 485.7mm
- Ug margin to limit
- 1.072mm
- Inverse-square time factor
- 1×
- Source half-life
- 1925.2d
- Decay-adjusted activity
- 925GBq
- Remaining activity
- 50%
- Activity correction
- 2×
Use the project-specified Section V edition and its native limit table. The exposure estimate assumes the same object, detector and beam conditions as the reference. Verify the approved technique and image-quality indicators. Equations and references